Test Code DHPVB Human Papillomavirus (HPV) Types 18, 31, 33, and 35, Droplet Digital PCR, Blood


Ordering Guidance


This test should be ordered to monitor circulating cell free DNA for the high-risk human papillomavirus (HPV)-18, -31, -33 or -35 genotypes in patients with a previous diagnosis. If the intent of the order is a pre-diagnostic screen for which the genotype is not previously known, order DHPVR / Human Papillomavirus (HPV) Type 16 with Type 18, 31, 33, and 35 Reflex, Droplet Digital PCR, Blood which can detect and quantify circulating cell free DNA from 5 high-risk human papillomavirus genotypes: HPV-16, -18, -31, -33 and -35.

 

For routine cervical cancer screening, order VHPV / Human Papillomavirus (HPV) Vaginal Detection with Genotyping for High-Risk Types by PCR.



Specimen Required


Supplies: Streck Tan Top Tube Kit (T715)

Container/Tube: Streck Cell-Free DNA blood collection kit

Specimen Volume: Two 10-mL Streck Cell-Free DNA blood collection tubes

Additional Information: Only blood collected in Streck Cell-Free DNA tubes will be accepted for analysis.


Secondary ID

622799

Useful For

Aid in the monitoring of cancers caused by human papillomavirus (HPV) high-risk genotypes 18, 31, 33, and 35

 

Disease monitoring after treatment of an HPV-related cancer and/or post-treatment surveillance as a prognostic indicator of cancer recurrence

 

This test should not be used to screen asymptomatic patients.

Method Name

Droplet Digital Polymerase Chain Reaction (ddPCR)

Reporting Name

HPV-18/31/33/35, cfDNA, ddPCR, B

Specimen Type

Whole blood

Specimen Minimum Volume

One 10 mL Streck Cell-Free DNA blood collection tube

Specimen Stability Information

Specimen Type Temperature Time
Whole blood Ambient (preferred) 7 days
  Refrigerated  7 days

Reject Due To

Specimen collected in tube other than Streck Cell-Free DNA collection tube Reject
  All specimens will be evaluated at Mayo Clinic Laboratories for test suitability.

Clinical Information

Human papillomavirus (HPV) is a double-stranded DNA virus from the Papillomaviridae family. There are over 200 known types of HPV, which can be classified into low-risk types, associated with benign conditions such as genital warts, and high-risk types, which are linked to the development of cancers, including cervical, anal, oropharyngeal, vulvar, penile, and vaginal cancers.(1-3)

 

Human papillomavirus infects the basal epithelial cells of the skin or mucous membranes. The virus integrates into host epithelial cells, where it can exist as either an episomal (extrachromosomal) or integrated form in the host DNA.(4,5) Low-risk HPV types (eg. HPV-6 and -11) usually remain in episomal form, while high-risk types (eg. HPV-16 and -18) are more likely to integrate into the host genome, causing dysregulation of the cell cycle. The virus produces proteins (E6 and E7) that interfere with tumor suppressor genes, p53 and pRb, and may lead to uncontrolled cell proliferation and a higher risk of progression to malignancy.(5)

 

Most HPV infections are transient and cleared by the host immune system within 1 to 2 years. However, persistent infection with high-risk HPV types increases the likelihood of progression to premalignant lesions and ultimately invasive cancer.(5)

 

Human papillomavirus cell free DNA has been shown to be an important biomarker circulating in patients with HPV-related cancers, including those of the head and neck. Fragmented HPV is released into circulation by apoptotic tumor cells and can be detected by molecular techniques. This assay utilizes droplet digital polymerase chain reaction to detect, subtype, and quantify high-risk HPV cfDNA in plasma samples.

Reference Values

Not detected

Interpretation

A qualitative result will be provided for each of the four high-risk human papillomavirus (HPV) genotypes (18, 31, 33 and 35) detected by this assay.

 

A result of "Not Detected" indicates the absence of HPV-18, -31, -33 and/or -35 cfDNA in the plasma (see Cautions below).

 

A result of "Detected" indicates the presence of HPV-18, -31, -33 and/or -35 cfDNA in the plasma. If HPV-18, -31, -33 and/or -35 cfDNA is detected, a quantitative value will be reported in fragments/mL.

 

The quantification range of this assay is 100 to 309,600 fragments/mL.

 

A result of "<100 fragments/mL" indicates that HPV-18, -31, -33 and/or -35 cfDNA is detected in the plasma, but the assay cannot accurately quantify the HPV-18, -31, -33 and/or -35 cfDNA present below this level.

 

A result of ">309,600 fragments/mL" indicates that HPV-18, -31, -33 and/or -35 cfDNA is detected in the plasma, but the assay cannot accurately quantify the HPV-18, -31, -33 and/or -35 cfDNA present above this level.

 

Samples with HPV signal below the limit of detection but above the limit of blank will be reported as "not detected" with an accompanying comment: "This specimen yielded signal above the limit of blank but below the assay's limit of detection, which may not be reproducible and should be interpreted in the context of the patient’s clinical presentation. If clinically indicated, recollect and submit a new specimen for testing in 2 to4 weeks."

Cautions

This assay only detects genetic fragments of the following HPV genotypes: 18, 31, 33 and 35.

 

This assay will not detect other human papillomavirus (HPV) genotypes.

 

This assay will not specify an anatomic location or source of detectable levels of HPV.

 

A result of “Not Detected” does not necessarily eliminate the possibility of HPV infection, or a tumor caused by HPV. Serial measurement of HPV in platelet poor plasma may be recommended based on clinical presentation.

 

This assay should only be used for patients with a clinical history and/or symptoms consistent with a confirmed or suspected HPV-related tumor.

 

Results must be interpreted in the context of the patient’s clinical picture.

 

Variants within the HPV E6 and E7 gene regions targeted by this assay may affect primer and/or probe binding, resulting in the under quantitation of viral fragments or failure to detect the presence of viral fragments.

Supportive Data

Analytical Sensitivity/Detection Capability:

The 95% limit of detection, by genotype, for this assay are:

19.45 fragments/mL of human papillomavirus (HPV)-18 cfDNA

17.83 fragments/mL of HPV-31 cfDNA

9.20 fragments/mL of HPV-33 cfDNA

7.71 fragments/mL of HPV-35 cfDNA

 

These were determined following a serial dilution of plasma spiked with a known concentration of each HPV genotype, with the limit of detection determined through probit analysis.

 

Accuracy:

Forty-eight contrived samples yielded 100% overall, positive and negative agreement.

 

Precision:

Intra- and inter-assay reproducibility experiments yielded 100% qualitative precision. Replicates at three levels within the analytical measuring range produced quantitative values with standard deviations of 0.00 – 0.07 log[10] fragments/mL.

 

Analytical Specificity:

A panel of 38 bacterial, fungal, parasitic and viral nucleic acid extracts from related organisms or organisms commonly found in plasma were tested by this assay with no cross-reactivity detected.

 

Inclusivity (in silico analysis) was performed using the NCBI GenBank database. Homology of assay primers and probes was determined using 25 representative strains from all four high-risk HPV genotypes detected by this assay (HPV-18, -31, -33, 35). Homology of oligos ranged from 88.9% to 100% across all strains analyzed.

 

Reference Range:

Platelet poor plasma from 20 healthy donors was tested by this assay. No samples yielded detectable HPV-18, -31, -33, nor -35 cfDNA above the limits of detection.

Clinical Reference

1. HPV and Cancer. National Cancer Institute. Updated May 9, 2025. Accessed October 14,2025. Available at www.cancer.gov/about-cancer/causes-prevention/risk/infectious-agents/hpv-and-cancer

2. Jamal Z, Anjum F. Oropharyngeal Squamous Cell Carcinoma. In: StatPearls [Internet]. Updated April 27, 2023. Accessed October 14, 2025. Available at www.ncbi.nlm.nih.gov/books/NBK563268/

3. Doorbar J, Quint W, Banks L, et al. The biology and life-cycle of human papillomaviruses. Vaccine. 2012;30 Suppl 5:F55–F70. doi:10.1016/j.vaccine.2012.06.083

4. Gribb JP, Wheelock JH, Park ES. Human papilloma virus (HPV) and the current state of oropharyngeal cancer prevention and treatment. Dela J Public Health. 2023;9(1):26-28. doi: 10.32481/djph.2023.04.008

5. Munger K, Phelps WC, Bubb V, Howley PM, Schlegel R. The E6 and E7 genes of the human papillomavirus type 16 together are necessary and sufficient for transformation of primary human keratinocytes. J Virol. 1989;63(10):4417-4421

6. Fernholz EC, Routman DM, Van Abel KM, et al. Detection, quantitation, and genotyping of human papillomavirus circulating tumor DNA by droplet digital PCR. J Clin Microbiol. 2025;63(9):e0058525. doi:10.1128/jcm.00585-25

Method Description

This laboratory-developed test utilizes whole blood samples collected in Streck Cell-Free DNA blood collection tubes. Cell-free DNA (cfDNA) is isolated from double-centrifuged plasma using the EZ2 Connect extraction system (Qiagen). Nucleic acid eluate is combined with polymerase chain reaction (PCR) reaction mix and partitioned into approximately 20,000 droplets by oil emulsion on the Automated Droplet Generator (Bio-Rad). Following PCR on the VeritiPro Thermal Cycler (Applied Biosystems), droplets are analyzed for presence or absence of amplified viral target sequences using the QX600 Droplet Reader (Bio-Rad).

 

The PCR reaction mix contains two primer and probe sets directed to the E6 and E7 gene regions of the human papillomavirus (HPV) type 18 genome, as well as single primer and probe sets for genotypes 31, 33 and 35. Amplicons range in size from 65 to 106 base pairs. The reaction mix also contains a primer-probe set directed to the human reference gene (ribonuclease P protein subunit 30 [RPP300]), which serves as an internal process control. Each hydrolysis probe is coupled with unique fluorescent dye that can be differentiated on the QX600. Each droplet is individually tallied and evaluated for fluorescence. Results are analyzed using dedicated software and Poisson statistics to provide absolute quantification of target and reference concentrations. Calculated results are reported as detected, not detected or indeterminate for HPV-18, -31, -33, and/or -35 cfDNA. A concentration is also reported for samples with detectable levels (fragments of HPV cfDNA per milliliter of plasma).(Unpublished Mayo method)

Day(s) Performed

Varies

Report Available

5 to 10 days

Specimen Retention Time

7 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

87799

LOINC Code Information

Test ID Test Order Name Order LOINC Value
DHPVB HPV-18/31/33/35, cfDNA, ddPCR, B In Process

 

Result ID Test Result Name Result LOINC Value
622876 HPV-18 ddPCR Result 61373-7
622877 HPV-18 ddPCR Quant In Process
622878 HPV-31 ddPCR Result 61375-2
622879 HPV-31 ddPCR Quant In Process
622880 HPV-33 ddPCR Result 61376-0
622881 HPV-33 ddPCR Quant In Process
622882 HPV-35 ddPCR Result 61377-8
622883 HPV-35 ddPCR Quant In Process